Experimental Study on in vivo Impedance Spectrums of Mice's Back Skin
Li Ying1, Song Bing2, Liu Zhipeng1, Yin Tao1#*
1(Institute of Biomedical Engineering, Chinese Academy of Medical Sciences & Peking Union Medical College, Tianjin 300192, China) 2(School of Dentistry, College of Biomedical and Life Sciences, Cardiff University, Cardiff CF144XY, UK)
Abstract:Different from other tissues, it is difficult to study the impedance of stratum corneum regarding skin as a whole subject, due to its huge impedance. The aim of this work is to non-invasively examine the characteristics of the impedance of mice’s back skin including stratum corneumin vivo. Using Agilent 4294A impedance analyzer, impedance spectrums of the back skin of 15 C57BL/6 mice under anesthesia were measured, and surface electrodes and acupuncture needles were used to record data as the whole skin and viable skin respectively. Along with the increase of frequency, the amplitude of impedance of the whole skin declined at the rate of -20 dB/dec within all the frequency span from 40 Hz to 107 Hz; whereas that of the viable skin decreased at -10 dB/dec in the range of 40 Hz to103 Hz, and at -3 dB/dec in the range of 103 Hz to 107 Hz. The phase of the whole skin decreased and then increased in a V-shape curve, and that of viable skin varied as a Π-shape curve. At low frequencies, such as 100 Hz, the whole skin impedance can be regarded as only consisting of the impedance of stratum corneum, and the viable skin part can be omitted. At medium frequencies, such as 105 Hz, the viable skin contributes over 10% to the whole skin impedance. However, at high frequencies, such as 107 Hz, the viable skin contributed over 80% to the whole skin impedance. The impedance of the whole skin and viable skin showed significant difference below 103 Hz (P<0.05) and no significant difference over 104 Hz. The results showed that this novel method that comparing the impedance recorded by surface electrodes and acupuncture needles was effective in the analysis of the impedance of stratum corneum within the whole skin impedance.
李 颖, 宋 冰, 刘志朋, 殷 涛. 小鼠背部皮肤的在体阻抗谱实验研究[J]. 中国生物医学工程学报, 2017, 36(2): 172-179.
Li Ying, Song Bing, Liu Zhipeng, Yin Tao. Experimental Study on in vivo Impedance Spectrums of Mice's Back Skin. Chinese Journal of Biomedical Engineering, 2017, 36(2): 172-179.
[1] 宋涛, 霍小林, 吴石增. 生物电磁特性及其应用[M]. 北京:北京工业大学出版社, 2008: 91-95. [2] Cranfield CG, Cornell BA, Grage SL, et al. Transient potential gradients and impedance measures of tethered bilayer lipid membranes: pore-forming peptide insertion and the effect of electroporation[J]. Biophys J, 2014,106(1): 182-189. [3] Fry CH, Salvage SC, Manazza A, et al. Cytoplasm resistivity of mammalian atrial myocardium determined by dielectrophoresis and impedance methods[J]. Biophys J, 2012,103(11): 2287-2294. [4] Di Biasio A, Ambrosone L, Cametti C. The dielectric behavior of nonspherical biological cell suspensions: an analytic approach[J]. Biophys J, 2010,99(1): 163-174. [5] Halter RJ, Hartov A, Poplack SP, et al. Real-time electrical impedance variations in women with and without breast cancer[J]. IEEE Trans Med Imaging, 2015,34(1): 38-48. [6] Aberg P, Nicander I, Hansson J, et al. Skin cancer identification using multifrequency electrical impedance - A potential screening tool[J]. IEEE Tran Biomed Eng, 2004,51(12): 2097-2102. [7] Keren H, Burkhoff D, Squara P. Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance[J]. Am J Physiol Heart Circ Physiol, 2007,293(1): H583-H589. [8] Yu Cheuk-Man, Wang Li, Chau E, et al. Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization[J]. Circulation, 2005,112(6): 841-848. [9] Corley A, Caruana LR, Barnett AG, et al. Oxygen delivery through high-flow nasal cannulae increase end-expiratory lung volume and reduce respiratory rate in post-cardiac surgical patients[J]. Br J Anaesth, 2011,107(6): 998-1004. [10] Vardavas CI, Anagnostopoulos N, Kougias M, et al. Short-term pulmonary effects of using an electronic cigarette: impact on respiratory flow resistance, impedance, and exhaled nitric oxide[J]. Chest, 2012,141(6): 1400-1406. [11] De Miguel-Etayo P, Moreno LA, Santabarbara J, et al. Anthropometric indices to assess body-fat changes during a multidisciplinary obesity treatment in adolescents: EVASYON Study[J]. Clin Nutr, 2015,34(3): 523-528. [12] Gatterer H, Schenk K, Laninschegg L, et al. Bioimpedance identifies body fluid loss after exercise in the heat: a pilot study with body cooling[J]. PLoS ONE, 2014,9(10): e109729. [13] Rosell J, Colominas J, Riu P, et al. Skin impedance from 1 Hz to 1 MHz[J]. IEEE Trans Biomed Eng, 1988,35(8): 649-651. [14] Gabriel C, Gabriel S, Corthout E. The dielectric properties of biological tissues: I. Literature survey[J]. Phys Med Biol, 1996,41(11): 2231-2249. [15] Martinsen G, Grimnes S, Haug E. Measuring depth depends on frequency in electrical skin impedance measurements[J]. Skin Research and Technology, 1999,5(3): 179-181. [16] Yamamoto T, Yamamoto Y. Electrical properties of the epidermal stratum corneum[J]. Med Biol Eng, 1976,14(2): 151-158. [17] Kalia YN, Pirot F, Guy RH. Homogeneous transport in a heterogeneous membrane: water diffusion across human stratum corneum in vivo[J]. Biophysical Journal, 1996, 71(5): 2692-2700. [18] Pethig R, Kell DB. The passive electrical-properties of biological-systems - their significance in physiology, biophysics and biotechnology[J]. Phys Med Biol, 1987,32(8): 933-970. [19] Schwan HP. Electrical properties of tissues and cell suspensions: mechanisms and models[C]// Proceedings of the 16th Annual International Conference of the IEEE EMBS. Baltimore: IEEE, 1994: A70-A71.