Application of Numerical Observer CHO in Evaluation of Filtering Method in PET
Xie Jing1, Yang Yong1*, Ye Hongwei2*, Chen Dongmei1
1School of Biomedical Engineering, Hangzhou Dianzi University, Hangzhou 310018, China 2Min Found Medical System Corporation Ltd., Shaoxing 312000, Zhejiang, China
Abstract:In clinical applications, it is necessary to limit the scan time and dose, which tends to lower the resolution of the positron emission tomography (PET) image and increase the noisein PET. A denoise method is required to achieve the clinically acceptable images, and a post filter after reconstruction is the most widely used method. Therefore, the comparison of smoothing effect of different filters, for instance the selection of filter parameters, is an important step in PET image reconstruction. Generally, the signal-to-noise ratio (SNR), recovery coefficient (RC) or similar methods are used in the parameter selections. But researchers still rely on their experience since those methods cannot be used quantitatively. As a generalized numerical observer, channelized hotelling observer (CHO) has been used in selection of various parameters in PET, such as reconstruction algorithm parameters, system design parameters, clinical protocol parameters and so on. However its application in the assessment of different filtering methods of image reconstruction is not widely studied. The purpose of this paper is to select the optimal parameters of two widely used filters, i.e. Gauss filter and Non-Local Mean (NLM) filter, and evaluate their smoothing effect in PET by comparing the area under the receiver operating characteristic(ROC)curve (AUC) values calculated by CHO. Experimental results showed that for the 13 mm sphere, Gauss filter with σ of1.1~1.4 and NLM filter with f of 0.5-0.9 achieved the maximum detectability, and for the 10 mm sphere, Gauss filter with σ of 1.4~2.0 and NLM filter with f of 0.5~0.9 achieved the maximum detectability. Though AUC values of both filters were as high as 0.9, the AUC value of NLM filter was larger than that of Gauss filter. It was also found out that bright spots had better contrast and lower noise in IEC images and patient images with the NLM filter than that with the Gauss filter. This conclusion was consistent with results obtained by traditional evaluation methods of the filter, which indicated that CHO accurately compared the performance of these two filters in the lesion detection task of PET.
谢靓, 杨勇, 叶宏伟, 陈冬梅. CHO数字观察器在PET图像滤波方法评估中的应用[J]. 中国生物医学工程学报, 2017, 36(6): 661-669.
Xie Jing, Yang Yong, Ye Hongwei, Chen Dongmei. Application of Numerical Observer CHO in Evaluation of Filtering Method in PET. Chinese Journal of Biomedical Engineering, 2017, 36(6): 661-669.
[1] Chan C, Meikle S, Fulton R, et al. A non-local post-filtering algorithm for PET incorporating anatomical knowledge[C]//2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC). Orlando: IEEE, 2009: 2728-2732. [2] Chun SY, Fessler JA, Dewaraja YK. Post-reconstruction non-local means filtering methods using CT side information for quantitative SPECT[J]. Physics in Medicine and Biology, 2013, 58(17): 6225-6240. [3] Xing Y, Hsiao IT, Gindi G. Rapid calculation of detectability in Bayesian single photon emission computed tomography.[J]. Physics in Medicine & Biology, 2003, 48(22):3755-3773. [4] Platia L, Goossens B, Vansteenkiste E, et al. Channelized Hotelling observers for the assessment of volumetric imaging data sets.[J]. Journal of the Optical Society of America, 2011, 28(6):1145-1163. [5] Brankov JG, Yang Y, Wei L, et al. Learning a channelized observer for image quality assessment[J]. IEEE Transactions on Medical Imaging, 2009, 28(7): 991-999. [6] Brankov JG. Evaluation of the channelized Hotelling observer with an internal-BV model in a train-test paradigm for cardiac SPECT defect detection[J]. Physics in Medicine and Biology, 2013, 58(20): 7159-7182. [7] Bonetto P, Qi J, Leahy RM. Covariance approximation for fast and accurate computation of channelized Hotelling observer statistics[J]. IEEE Transactions on Nuclear Science, 2000, 47(4):1567-1572. [8] Kinahan PE, Kim JS, Lartizien C, et al. A comparison of planar versus volumetric numerical observers for detection task performance in whole-body PET imaging[J]. IEEE Transactions on Nuclear Science, 2002, 51(1):34-40. [9] El FG, Surti S, Trott CM, et al. Improvement in lesion detection with whole-body oncologic time-of-flight PET.[J]. Journal of Nuclear Medicine, 2011, 52(3):347-353. [10] Wunderlich A, Noo F. New theoretical results on channelized Hotellingobserverperformance estimation with known difference of class means[J]. IEEE Transactions on Nuclear Science, 2013, 60(1):182-193. [11] Lartizien C, Buvat I. Comparison of 3D PET data bootstrap resampling methods for numerical observer studies[C]//2005 IEEE Nuclear Science Symposium Conference Record. Fajardo: IEEE, 2005: 2138-2147. [12] Qi J, Huesman RH. Lesion detection and quantitation of positron emission mammography[C]//2001 IEEE Nuclear Science Symposium Conference Record. San Diego: IEEE, 2001: 2248-2252. [13] Janeiro L, Comtat C, Lartizien C, et al. Numerical observer studies comparing FORE+ AWOSEM, FORE+ NECOSEM and NEC based fully 3-D OSEM for 3-D whole-body PET imaging[J]. IEEE Transactions on Nuclear Science, 2006, 53(3): 1194-1199. [14] NEMA NU 2-2012,Performance Measurements of Positron Emission Tomographs [S]. [15] 李红艳. PET 图像重建算法研究及实现[D]. 长沙:中南大学, 2009. [16] Wunderlich A, Noo F, Gallas BD, et al. Exact confidence intervals for channelized Hotelling observer performance in image quality studies[J]. IEEE Transactions on Medical Imaging, 2015, 34(2): 453-464. [17] 林桂粉, 罗立民. 基于三维高斯滤波去噪加速蒙特卡罗剂量分布模拟[J]. 中华放射医学与防护杂志, 2006, 26(4):389-390. [18] Dutta J, Leahy RM, Li Q. Non-local means denoising of dynamic PET images[J]. PLoS ONE, 2013, 8(12): e81390. [19] Chun SY, Fessler JA, Dewaraja YK. Non-local means methods using CT side information for I-131 SPECT image reconstruction[C]//2012 Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). Anaheim: IEEE, 2012: 3362-3366. [20] Surti S, Karp S, Popescu LM, et al. Investigation of time-of-flight benefit for fully 3-DPET[J]. IEEE Transactions on Medical Imaging, 2006, 25(5):529-538. [21] Surti S, Karp J S. Experimental evaluation of a simple lesion detection task with time-of-flight PET[J]. Physics in Medicine and Biology, 2008, 54(2): 373-384. [22] 边兆英, 黄静, 马建华,等. 投影数据恢复导引的PET图像滤波[J]. 计算机工程与应用, 2011, 47(17):196-199. [23] Jomaa H, Mabrouk R, Morain NF, et al. Multi-scale and non local mean based filter for positron emission tomography imaging denoising[C]// International Conference on Advanced Technologies for Signal and Image Processing. Monastir: IEEE, 2016:108-112.