|
|
Bone Healing Analysis of Transverse and Oblique Fracture Under Different Internal Fixation Parameters |
Fang Runxin1, Ji Aimin1*, Chen Changsheng1,2, Long Dengyan1, Zhao Zhonghang1 |
1(College of Mechanical and Electrical Engineering,Hehai University, Changzhou, Jiangsu 213022, China) 2(Changzhou Orthmed Medical Instrument Co., Ltd., Changzhou, Jiangsu 213000, China) |
|
|
Abstract The aim of this work was to study the influences of internal fixation parameters on the fracture healing between transverse and oblique fracture. The second-development of Abaqus based on Python was used for the simulation of healing process. During the simulating process, the mechanical modulation and cell differentiation were integrated. The method of orthogonal experiment was used for studying the effects of parameters on the healing process, and different parameter combinations were constructed based on the orthogonal table of L9(34). Finally, an optimal combination was selected after the analysis. Results showed that the working length of the plate was the most influential parameter, a great divergence of healing pattern between transverse and oblique fracture was found under the low level of working length, the time of healing reached 1 000 MPa for transverse fracture was shorter than that for oblique fracture. Increased working length will drive the two fracture modes to show a similar healing pattern, their healing results finally reached or over 1400 MPa at the end of 16 weeks, which meant that a suitable mechanical environment for healing process could be ensured by omitting one or two screws that were proximal to the fracture gap. The healing efficiency of both the transverse and oblique fracture was highly correlated with the choice of internal fixation parameters. Compared with transverse fracture, oblique fracture required higher standard of stability for fracture healing, therefore, more screws were needed for the oblique fracture to ensure the stability.
|
Received: 13 March 2018
|
|
|
|
|
[1] 张魁忠, 吴柄华, 徐聪,等. 螺钉数量对钢板螺钉内固定结构载荷的生物力学影响[J]. 中国组织工程研究, 2009, 13(43):8454-8457. [2] Field JR, Törnkvist H, Hearn TC, et al. The influence of screw omission on construction stiffness and bone surface strain in the application of bone plates to cadaveric bone [J]. Injury-International Journal of the Care of the Injured, 1999, 30(9):591-598. [3] Törnkvist H, Hearn TC, Schatzker J. The strength of plate fixation in relation to the number and spacing of bone screws [J]. Journal of Orthopaedic Trauma, 1996, 10(3):204-208. [4] 盛伟, 纪爱敏, 陈长胜. 基于有限元分析的股骨干骨折内固定系统多因素的正交试验设计[J]. 生物医学工程研究, 2017, 36(2):146-151. [5] 林振恩, 谢丹, 张森. 股骨远端骨折锁定钢板螺钉布局的有限元分析[J]. 中国组织工程研究, 2014, 18(22):3517-3522. [6] 刘雄, 吴卫东, 荀福兴,等. 不同长度锁钉固定桡骨远端不稳定骨折的生物力学比较[J]. 医用生物力学, 2014, 29 (2):161-166. [7] 张先龙, 戴克戎, 汤亭亭. 应力松弛接骨板对骨折愈合影响的生物力学研究[J]. 中国骨与关节损伤杂志, 2000, 15(1):30-32. [8] 程斌, 丁真奇, 姚小涛,等. 轴向应力促进胫骨骨折愈合的三维有限元力学参数优化研究[J]. 中国骨与关节损伤杂志, 2016, 31(8):839-842. [9] 曲爱丽, 王冬梅, 景捷. 骨折愈合进程下髁突囊内骨折内固定的有限元分析[J]. 中国生物医学工程学报, 2015, 34(4):507-512. [10] Mehboob A, Mehboob H, Chang SH, et al. Effect of composite intramedullary nails (IM) on healing of long bone fractures by means of reamed and unreamed methods[J]. Composite Structures, 2017, 167:76-87. [11] Isaksson H, Wilson W, van Donkelaar CC, et al. Comparison of biophysical stimuli for mechano-regulation of tissue differentiation during fracture healing [J]. Journal of Biomechanics, 2006, 39(8):1507-1516. [12] Mehboob H, Chang SH. Effect of structural stiffness of composite bone plate–scaffold assembly on tibial fracture with large fracture gap [J]. Composite Structures, 2015, 124:327-336. [13] Son DS, Mehboob H, Chang SH. Simulation of the bone healing process of fractured long bones applied with a composite bone plate with consideration of the blood vessel growth[J]. Composites Part B Engineering, 2014, 58(3):443-450. [14] Lacroix D, Prendergast PJ, Li G, et al. Biomechanical model to simulate tissue differentiation and bone regeneration: application to fracture healing.[J]. Medical & Biological Engineering & Computing, 2002, 40(1):14-21. [15] Mehboob A, Mehboob H, Kim J, et al. Influence of initial biomechanical environment provided by fibrous composite intramedullary nails on bone fracture healing[J]. Composite Structures, 2017, 175:123-134. [16] Mehboob H, Chang SH. Optimal design of a functionally graded biodegradable composite bone plate by using the Taguchi method and finite element analysis[J]. Composite Structures, 2015, 119:166-173. [17] 方斌, 杨海韵. 骨折功能锻炼的生物力学研究进展[J]. 中国骨伤, 2009, 22(6):478-480. [18] Rozbruch SR, Müller U, Gautier E, et al. The evolution of femoral shaft plating technique[J]. Clinical Orthopaedics & Related Research, 1998, 354(354):195-208. [19] Gómezbenito MJ, Garcíaaznar JM, Kuiper JH, et al. Influence of fracture gap size on the pattern of long bone healing: a computational study[J]. Journal of Theoretical Biology, 2005, 235(1):105-119. [20] Jepsen KJ, Davy DT, Krzypow DJ. The role of the lamellar interface during torsional yielding of human cortical bone[J]. Journal of Biomechanics, 1999, 32(3):303-310. [21] 张建新, 王和鸣, 陈日齐,等. 三维有限元法分析横形斜形骨折对骨折愈合的影响[J]. 医用生物力学, 2005, 20(4):226-230. [22] 马克昌. 骨生理学[M]. 郑州:河南医科大学出版社, 2000. [23] 张建新, 王和鸣, 陈日齐,等. 从生物力学角度探讨斜形与横形骨折对骨折愈合的影响[J]. 中国骨伤, 2002, 15(11):671-673. [24] Miramini S, Zhang L, Richardson M, et al. Influence of fracture geometry on bone healing under locking plate fixations: A comparison between oblique and transverse tibial fractures[J]. Medical Engineering & Physics, 2016, 38(10):1100-1108. [25] Stoffel K, Dieter U, Stachowiak G, et al. Biomechanical testing of the LCP-how can stability in locked internal fixators be controlled?[J]. Injury-international Journal of the Care of the Injured, 2003, 34(Suppl 2):11-19. [26] 胡新佳, 杨述华, 林博文,等. 胫骨中段骨折螺钉数量的生物力学实验[J]. 中国组织工程研究, 2009, 13(17):3223-3227. |
|
|
|