Analysis Method of Current Density Based on Magnetocardiosignal
Ai Haiming1#, Primin Mykhailo2, Mi Wang2, Wu Hongjin3*
1(Faculty of Science and Technology, Beijing Open University, Beijing 100081, China) 2(Beijing Milestone Science & Technology Development Co.,Ltd, Beijing 101500, China) 3(Beijing Haidian Section of Peking University Third Hospital, Beijing 100080, China)
Abstract:The analysis of current density vector as diagnostic tool is a critical technique in clinical application of magnetocardiography, its reconstruction and classification performance will directly determine the accuracy of cardiac disease screening. A current density analysis method was studied in this paper, which includes inverse problem solutionfor current density reconstruction and new proposed algorithm of automatic classification. The current density analysis method was used to clinically validate 27 healthy volunteers and 75 patients with myocardial ischemia. First, the SQUID sensor of four-channel magnetocardiography was used to acquire magnetocardiosignal in 36 measurement sites, ECG signal base on standard lead ⅡECG was applied to synchronize magnetocardiosignal. Second, inverse problem solution algorithm based on Fourier transformation was applied to reconstruct current density distribution for each subject, automatic classification parameters of CDI, IPCD, ERT were automatically calculated, the linear discriminant function of healthy group and myocardial ischemia group in automatic classification algorithm was determined by multiple linear regression model. Finally, automatic classification of reconstructed current density was performed by the new automatic classification algorithm based on KL entropy calculation and liner discriminant function. Results showed that significant difference of three classification parameters of CDI, IPCD, ERT between healthy group and myocardial ischemia group was 2.0×10-11, 2.7×10-5 and 3.0×10-4, respectively; sensitivity and specificity in myocardial ischemia diagnosis for CDI, IPCD, ERT was 73%, 81%, 60% and 100%, 65%, 73%, respectively; sensitivity, true positive rate, specificity, false positive rate and true negative rate of automatic classification results was 89.33%, 100%, 100%, 0% and 76.47%, respectively. Therefore, the proposed current density analysis method is reliable, accurate and available for clinical diagnosis of myocardial ischemia.
艾海明, Primin Mykhailo, 米旺, 吴红金. 基于心磁信号的电流密度分析方法研究[J]. 中国生物医学工程学报, 2020, 39(2): 170-179.
Ai Haiming, Primin Mykhailo, Mi Wang, Wu Hongjin. Analysis Method of Current Density Based on Magnetocardiosignal. Chinese Journal of Biomedical Engineering, 2020, 39(2): 170-179.
[1] Her AY, Park JW. Repolarization heterogeneity of magnetocardiography predicts long-term prognosis in patients with acute myocardial infarction[J]. Yonsei Med, 2016, 57(6): 1305-1306.
[2] Cui GY, Uchikawa Y, Kotan M. Three dimensional measurement of MCG in normal subject[J]. Chinese Journal of Biomedical Engineering(English Edition), 1999, 2: 9-16.
[3] Ma J, Uchiyama T. Development of peak-to-peak voltage detector- type MI Gradiometer for magnetocariography[J]. IEEE Transactions on Magnetics, 2018, 54(11): 1-5.
[4] Zhang Shulin, Zhang Guofeng, Wang Yongliang, et al. Multichannel fetal magnetocardiography using SQUID bootstrap circuit [J]. Chinese Physics B, 2013, 22(12):514-516.
[5] Luiz B, Wayne G, Sridharam R, et al. Cardiac late Na+ current: proarrhythmic effects, roles in long QT syndromes, and pathological relationship to CaMKII and oxidative stress [J]. Heart Rhythm, 2015, 12(2): 440-443.
[6] Timothy OL, Alex HW, Alessio F, et al. Cell types, network homeostasis and pathological compensation from a biologically plausible ion channel expression model [J]. Neuron, 2014, 82(4): 809-821.
[7] Peeyush S, Rahul K, Emmanuel S, et al. Magentocardiography-derived 3D current density maps show increased sensitivity and localization of ischemic damage in high risk unstable angina patients [J]. Circulation, 2016, 134(1): 182-189.
[8] Vasnev AV, Maslennikov YV, Primin MA, et al. Magnetocardiographic noninvasive imaging of the right artium stretchability due to short-term overload during respiratory maneuvers [C]// World congress on Medical Physics and Biomedical Engineering. Munich: Springer Link, 2009: 668-670.
[9] Comani S, Romani GL, Silvano DL, et al. Magnetocardiography: current status and perspective. Part 1: principles and instrumentation [J]. Ital Heart J, 2002,3:75-85.
[10] Primin MA, Nedaivada IV, Maslennikov YV, et al. Software for the magnetocardiographic complex for the early diagnostics and monitoring of heart deceases[J]. Journal of Communications Technology and Electronics, 2010, 55(10):1169-1186.
[11] Primin MA, Nedaivada IV. A mathematical model and measurements algorithms for a dipole source location [J]. Int Journal of Applied Electromagnetics and Mechanics, 1997, 8: 119-131.
[12] 孙慧娜, 唐发宽, 黄骁, 等. 心磁图的主要临床应用及研究进展[J]. 中国循证心血管医学杂志,2014, 6(4): 499-508.
[13] Chang Yicheng, Wu Chauchung, Lin Chihung, et al. Early myocardial repolarization heterogeneity is detected by magnetocardiography in diabetic patients with cardiovascular risk factors [J]. PLoS ONE, 2015,1-12.
[14] Roberta G, Tracey C, Joseph M, et al. Review on solving the inverse problem in EEG source analysis [J]. Journal of NeuroEngineering and Rehabilitation,2008, 5(25): 1-33.
[15] Dossel O. Inverse problem of electro- and magnetocardiography: review and recent progress [J]. International Journal of Bioelectro- magnetism, 2000, 2(2): 1-56.
[16] Yaman F, Yakho VG, Potthast R. A survey on inverse problems for applied sciences [J]. Mathematical Problems in Engineering, 2013, 2013:1-19.
[17] Ioannidest AA, Boltont JRP, Clarket CJS. Continuous probabilistic solutions to the biomagnetic inverse problem [J]. Inverse Problem, 1990,6(4):523-542.
[18] Iwaki S, Shoogo U. Weighted minimum-norm source estimation of magnetoencephalography utilizing the temporal information of the measured data [J]. Journal of Applied Physics, 1998, 11(83): 6441-6443.
[19] Zuber JW, Wells FS, Fedoseev SA, et al. A new approach to the inverse problem for current mapping in thin-film superconductors [J]. Journal of Applied Physics, 2018, 123: 1-10.
[20] Primin MA, Nedayvoda IV. A mathematical model and measurement algorithms for a dipole source location [J]. International Journal of Applied Electromagnetics and Mechanics, 1997, 8: 119-131.
[21] 朱红毅. 基于脑磁图的电流偶极子反演研究 [D]. 杭州:浙江大学,2005.
[22] Bang WD, Kim K, Lee YH, et al. Repolarization heterogeneity of magnetocardiography predicts long-term prognosis in patients with acute myocardial infarction [J]. 2016, Yonsei Med J, 2016, 57(6): 1339-1346.
[23] Lim HK, Kwon H, Chung N, et al. Usefulness of magnetocardiogram to detect unstable angina pectoris and non-ST elevation myocardial infarction [J]. The American Journal of Cardiology, 2009, 103: 448-454.
[24] Primin MA, Nedayvoda IV. Inverse problem solution algorithms in magnetocardiography: new analytical approaches and some results [J]. International Journal of Applied Electromagnetic and Mechanics, 2009, 29: 65-81.
[25] Primin MA, Nedayvoda IV. A method and an algorithm to reconstruct the spatial structure of current density vectors in magnetocardiography [J]. Cybernetics and Systems Analysis, 2017,53(3):485-494.
[26] Fagaly RL. Superconducting quantum interference device instruments and applications [J]. Review of Scientific Instruments,2006,77(101101):1-45.
[27] Ivanov GG, Bulanova NA, Vostrikov VA, et al. Automatic classification of patients groups using magnetocardiography in the diagnosis of the atria lesions in patients with chronic obstructive pulmonary disease and coronary heart disease (in Russian) [J]. Rudn Journal of Medicine, 2015, 2:62-72.
[28] John JW, Olle P, James WW, et al. Criteria for ECG detection of acute myocardial ischemia: sensitivity versus specificity [J]. 2018, 51(6): S12-S17.
[29] Akihiko K, Kuniomi O, Yasushi W, et al. space-time database for standardization of adult magnetocardiogram-making standard MCG parameters [J]. Pace,2008,31:422-431.
[30] Toistrup K, madsen B, Ruiz J, et al. Non-invasive resting magnetocardiographic imagning for the rapid detection of ischemia in subjects presenting with chest pain [J]. Cardiology, 2006, 106: 270-276.
[31] Li Yingmei, Che Zaiqian, Quan Weiwei, et al. Diagnostic outcomes of magneocardiography in patients wth coronary artery disease [J]. Int J Clin Exp Med, 2015, 8(2): 2441-2446.
[32] Chaikovsky I, Sydorova NN. Coronary artery diseases [M]. Rijeka: Intech Open Pub, 2012: 70-73.