Abstract:In this paper, a synthetic algorithm was proposed to extract the fetal electrocardiogram (FECG) from single-lead abdominal signals. We extracted maternal electrocardiogram (MECG) and FECG successively and then calculated maternal and fetal heart rate. In the algorithm, we first applied Teager energy operator with parameter k (k=19) to protrude QRS waves of MECG, so that the location of maternal R peaks was detected correctly by simple threshold. Then we resampled between every adjacent R peaks to get same R-R interval T. After applying a comb filter whose teeth coincide with the maternal R-R interval T, we obtained the MECG with same R-R interval. And the real MECG was obtained by resampling between every adjacent R peaks in the filtered signals again to recover the original R-R intervals. The extracted MECG was subtracted from the abdominal signals. In the residual signal the QRS waves of FECG had better signal-noise-ratio. Next, the same procedure was applied on the residual signal to extract the FECG signal. We analyzed 8 sets of abdominal signals (total 26 channels) that were selected from the Physionet non-invasive FECG database. The sensitivity (Se), positive predictive value (PPV) and accuracy (F1) of fetal QRS waves detection were calculated. Results showed that the detection accuracy of fetal QRS waves reached 87.1%, with six channel even reached 100%. The maternal heart rate (MHR) and fetal heart rate (FHR) of each channel were also calculated. We found out that the MHRs, as well as the FHRs, showed good consistence with channels of the same set. In the same set, the maximum error of the mean MHR was 0.1 times per minute and the maximum error of mean FHR was 0.9 times per minute. Such a phenomenon also proved the reliability of the proposed algorithm.
[1] Deans AC, Steer PJ. The use of the fetal electrocardiogram in labour[J]. Br J Obstetrics Gynaecol,1994,101:9-17. [2] Ferrara ER, Widrow B. Fetal electrocardiogram enhancement by time sequenced adaptive filter[J]. IEEE Trans Biomed Eng, 1982, 29:458-460. [3] Martinez M, Soria E, Calpe J, et al. Application of the adaptive impulse correlated filter for recovering fetal electrocardiogram[J]. Computers in Cardiology, 1997, 24: 9-12. [4] Mooney DM, Grooome LJ, Bentz LS, et al. Computer algorithm for adaptive extraction of fetal cardiac electrical signal[C]//ACM Symp Appl Comput, Nashville: ACM, 1995: 113-117. [5] Ungureanu M, Bergmans JWM, Oei SG, et al. Fetal ECG extraction during labor using an adaptive maternal beat subtraction technique[J]. Biomedizinische Technik, 2007, 52(1):56-60 [6] Sheikh Algunaidi MM, Mohd Ali MA, Fokhrul Islam M. Evaluation of an improved algorithm for fetal QRS detection[J]. International Journal of the Physical Sciences, 2011,6(2): 213-220. [7] Zarzoso V, Nandi AK, Bacharakis E. Maternal and fetal ECG separation using blind source separation methods[J]. IMA J Math Appl Med Biol, 1997,14:207-255. [8] Immanuel JRJ, Prabhu V, Chrstopheraj JV, et al. Separation of maternal and fetal ECG signal from the mixed source signal using FASTICA[J]. Procedia Eng, 2012,30:356-363. [9] Camps Valls G, Martinez Sober M, Soria Olivas E, et al. Fetal ECG recovery using dynamic neural networks[J]. Artif Intell Med, 2004,31:197-209. [10] Assaleh K. Extraction of fetal electrocardiogram using adaptive neuro-fuzzy inference systems[J]. IEEE Trans Biomed Eng, 2007,54(6):59-68. [11] Kanjilal PP, Palit S, Saha G. Fetal ECG extraction from single-channel maternal ECG using singular value decomposition[J]. IEEE Trans Biomed Eng, 1997,44(1):51-59. [12] Khamene A, Negahdaripour S. A new method for the extraction of fetal ECG from the composite abdominal signal[J]. IEEE Trans Biomed Eng, 2000,47(4):507-516. [13] Mochimaru F, Fujimoto Y, Ishikawa Y. Detecting the fetal electrocardiogram by wavelet theory-based methods[J]. Prog Biomed Res, 2002,7(3):185-93. [14] Huang NE, Shen Zheng, Long SR, et al. The empirical mode decomposition and the Hilbert spectrum for nonlinear and non-stationary time series analysis[J]. Proc R Soc Lond A, 1998, 454(1971):904-993. [15] Liu Guangchen, Luan Yihui. An adaptive integrated algorithm for noninvasive fetal ECG separation and noise reduction based on ICA-EEMD-WS[J]. Med Biol Eng Comput, 2015,53:1113-1127. [16] Zheng Wei, Liu Hongxing, He Aijun, et al. Single-lead fetal eclectrocardiogram estimation by means of combining R-peak detection, resampling and comb filter[J]. Medical Engineering & Physics, 2010,32:708-719. [17] Choi JH, Jung HK, Kim T. A new action potential detector using the MTEO and its effects on spike sorting system at low signal-to-noise ratios[J]. IEEE Trans Biomed Eng, 2006,53(4):738-746. [18] Maragos P, Kaiser JF, Quatieri TF. On amplitude and frequency demodulation using energy operators[J]. IEEE Trans Signal Process, 1998,41(4):180-187. [19] Guerrero JM, Sober MM, Bataller MM. New algorithm for fetal QRS detection in surface abdominal records[J]. Computation in Cardiology, 2006, 33:441-444. [20] Behar J, Johnson A, Glifford GD, et al. A comparison of single channel fetal ECG extraction methods[J]. Annals of Biomedical Engineering, 2014,42(6):1340-1353.