Particle Image Velocimetry Experimental Study on the Flow Field in the Outflow Cannula of an Axial Left Ventricular Assist Pump |
1 Chinese Academy of Medical Sciences and Peking Union Medical College, Fuwai Hospital, National Center for Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, Beijing 100037, China
2 Department of Cardiovascular Surgery, Chinese Academy of Medical Sciences and Peking Union Medical College, Fuwai Hospital, Beijing 100037, China |
|
|
Abstract The aim of this work is to study the blood flow distribution in outflow cannula of an axial left ventricular pump and analyze the risk of thrombosis in the axial blood pump outflow cannula according to the relationship between blood flow property and thrombogenesis. We obtained the flow distribution in the central section of the axial blood pump outflow cannula with the two dimensional particle image velocimetry (PIV) system. Then the blood flow distribution in the whole outflow cannula with the three dimensional PIV system was obtained. The rotate speed of the blood pump was set to (10000±20) r/min with flux of 8.05 L/min in the testing. The possibility of thrombogenesis in the axial blood pump outflow cannula was estimated through analyzing the flow property in the outflow cannula. There were spiral flow and flow normal to the wall of cannula nearby the outlet of the blood pump. The two kinds of flow weaken off when the blood flowed through the outflow cannula, so the blood flow far away from the pump outlet was steady. There was not backflow, vortex flow or lowspeed flow region in the whole cannula. The blood flow velocity along the cannula wall increased to the maximum 1.0 m/s from 0 m/s rapidly outside the boundary layer. The blood speed range reduced from 0.7 m/s~2.2 m/s to 1.0 m/s~1.5 m/s along the cannula. The turbulence level near the pump outlet is 0.31. The turbulence level far away from the pump outlet is 0.15. The flow in the cannula become smooth and steady gradually. It is difficult to form thrombus in the cannula caused by blood flow property because there was no backflow, vortex flow or stagnant flow region in the cannula. The outflow cannula can decrease the spiral flow, which is helpful to improve the flow characteristics of the blood flowing from the axial blood pump. The improved flow characteristics can decrease the injury of aorta caused by disorder blood flow.
|
|
|
|
|
[1]John R, Naka Y, Smedira NG, et al. Continuous flow left ventricular assist device outcomes in commercial use compared with the prior clinical trial [J]. Ann Thorac Surg, 2011, 92(4): 1406-1413.
[2]Kirklin JK, Naftel DC, Pagani FD, et al. Longterm mechanical circulatory support (destination therapy): on track to compete with heart transplantation? [J] J Thorac Cardiovasc Surg, 2012, 144(3): 584-603.
[3]Rodriguez LE, Suarez EE, Loebe M, et al. Ventricular assist devices (vad) therapy: new technology, new hope?[J]. Methodist Debakey Cardiovasc J, 2013, 9(1): 32-37.
[4]Shah NR, Cevik C, Hernandez A, et al. Transthoracic echocardiography of the HeartWare left ventricular assist device [J]. J Card Fail, 2012, 18(9): 745-748.
[5]Swiecicki PL, Edwards BS, Kushwaha SS, et al. Left ventricular device implantation for advanced cardiac amyloidosis [J]. J Heart Lung Transplant, 2013, 32(5): 563-568.
[6]Muthiah K, Phan J, Robson D, et al. Centrifugal continuousflow left ventricular assist device in patients with hypertrophic cardiomyopathy: a case series [J]. ASAIO J, 2013, 59(2): 183-187.
[7]Abicht T, Gordon R, Meehan K, et al. Complex HeartMate II infection treated with pump exchange to HeartWare HVAD [J]. ASAIO J, 2013, 59(2): 188-192.
[8]Poullis M. Computational fluid dynamic analysis to prevent aortic root and valve clots during left ventricular assist device support [J]. Journal of ExtraCorporeal Technology, 2012, 44(4): 210-215.
[9]Kosaka R, Nishida M, Maruyama O, et al. Development of a miniaturized massflow meter for an axial flow blood pump based on computational analysis [J]. J Artif Organs, 2011, 14(3): 178-184.
[10]Karmonik C, Partovi S, Loebe M, et al. Influence of LVAD cannula outflow tract location on hemodynamics in the ascending aorta: a patientspecific computational fluid dynamics approach [J]. ASAIO J, 2012, 58(6): 562-567.
[11]Su Boyang, Chua LP, Lim TM, et al. Evaluation of the impeller shroud performance of an axial flow ventricular assist device using computational fluid dynamics [J]. Artif Organs, 2010, 34(9): 745-759.
[12]蔺嫦燕,侯晓彤,吴广辉,等. XZII型轴流血泵的流场分析 [J]. 北京生物医学工程, 2005, 24(6): 405-409.
[13]陈建中,张锡文,何枫,等. 微型轴流血泵后导叶对流场的影响研究 [J]. 北京生物医学工程, 2007, 26(1): 1-5.
[14]Triep M, Brucker C, Schroder W, et al. Computational fluid dynamics and digital particle image velocimetry study of the flow through an optimized microaxial blood pump [J]. Artif Organs, 2006, 30(5): 384-391.
[15]Untaroiu A, Wood HG, Allaire PE, et al. Computational design and experimental testing of a novel axial flow LVAD [J]. ASAIO J, 2005, 51(6): 702-710.
[16]Yang Xiaochen, Zhang Yan, Gui Xingmin, et al. Computational fluid dynamicsbased hydraulic and hemolytic analyses of a novel left ventricular assist blood pump [J]. Artif Organs, 2011, 35(10): 948-955.
[17]Throckmorton AL, Lim DS, McCulloch MA, et al. Computational design and experimental performance testing of an axialflow pediatric ventricular assist device [J]. ASAIO J, 2005, 51(5): 629-635.
[18]Tuzun E, Narin C, Gregoric ID, et al. Ventricular assist device outflowgraft site: effect on myocardial blood flow [J]. J Surg Res, 2010, 171(1): 71-75.
[19]Manning KB, Miller GE. Flow through an outlet cannula of a rotary ventricular assist device [J]. Artif Organs, 2002, 26(8): 714-723.
[20]Song Xinwei, Wood HG, Allaire PE, et al. Inlet and outlet devices for rotary blood pumps [J]. Artif Organs, 2004, 28(10): 911-915.
[21]Yang Xiaochen, Gui Xingmin, Huang Hui, et al. Particle image velocimetry experimental and computational investigation of a blood pump [J]. Journal of Thermal Sience, 2012, 21(3): 262-268.
[22]Day SW, McDaniel JC. PIV measurements of flow in a centrifugal blood pump: timevarying flow [J]. J Biomech Eng, 2005, 127(2): 254-263.
23]王芳群,李岚,钱坤喜,等. 基于粒子图像测速技术的血泵内流场研究 [J]. 中国生物医学工程学报, 2009, 28(4): 597-601.
[24]Akagawa E, Lee H, Tatsumi E, et al. Flow visualization for different port angles of a pulsatile ventricular assist device [J]. J Artif Organs, 2011, 15(2): 119-127.
[25]韩占忠,王敬,兰小平. FLUENT流体工程仿真计算实例与应用 [M]. 北京: 北京理工大学出版社, 2007: 9-10. |
|
|
|