中国生物医学工程学报  2018, Vol. 37 Issue (4): 429-434    DOI: 10.3969/j.issn.0258-8021.2018.04.006
  论著 本期目录 | 过刊浏览 | 高级检索 |
基于S-G滤波的超声造影分析参数灌注成像
王本刚1,2, 丁红1*, 彭诗云1, 许智婷1,2, 付甜甜1,2, 王文平1
1(复旦大学附属中山医院超声科,上海 200032)
2(上海市影像医学研究所,上海 200032)
Analysis of Parameter Perfusion Imaging for Contrast-Enhanced Ultrasound Based on S-G Filter
Wang Bengang1,2, Ding Hong1*, Peng Shiyun1, Xu Zhiting1,2, Fu Tiantian1,2, Wang Wenping1
1(Department of Ultrasound, Zhongshan Hospital, Fudan University, Shanghai 200032, China)
2(Shanghai Institute of Medical Imaging, Shanghai 200032, China)
全文: PDF (16493 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 为弥补以往超声造影定量分析软件将感兴趣区(ROI)作为均质的整体区域进行分析的不足,提出一种基于S-G滤波器的在时域内直接处理数据平滑问题的定量分析方法,针对ROI区域内部像素点进行灌注成像参数分析,快速并可最大限度地保留原始有用信息。首先获得超声造影的原始DICOM动态医学数据,获取ROI内每个像素点的时间强度曲线,用S-G滤波器对时间强度曲线进行分析,得到反映灌注时相和灌注强度的多项定量分析参数,如到达时间(AT)、达峰时间(TTP)、达峰强度(PI)、平均通过时间(MTT)、曲线下面积(AUC)和上升斜率(RS)等;然后再对像素点参数分别进行彩色编码并以不同颜色显示,从而准确量化和直观显示超声造影的各项灌注参数。为评估该分析方法的稳定性,比较3位研究者分别对15个肝细胞癌病灶的分析结果。经统计,所有参数的组间相关系数(ICC)>0.80,验证了该方法的可行性和稳定性。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
王本刚
丁红
彭诗云
许智婷
付甜甜
王文平
关键词 超声造影时间强度曲线拟合分析参数灌注成像S-G滤波    
Abstract:The previous perfusion analysis software for contrast-enhanced ultrasound (CEUS) provided quantitative parameters under the assumption that the region of interest (ROI) was a whole homogeneous perfusion area. To overcome the imperfection, we herein presented a fast data smoothing method with time domain based on S-G filter of ROI’s internal regional perfusion imaging based on pixel parameter analysis. The time intensity curve (TIC) of each pixel was obtained from the original DICOM dynamic imaging of CEUS data, the quantitative analysis of parameters of TIC were conducted based on the S-G filter, including time of arrival (AT), time to peak (TTP), peak intensity (PI), mean transit time (MTT), the area under the curve (AUC), and rising slope (RS). After that the pixel perfusion parameters coding was displayed in the color mode accurately and intuitively. The stability test was performed on 15 cases of hepatocellular carcinomas to obtain different quantitative parameters of TIC by three investigators respectively. The intraclass correlation coefficient of all parameters was more than 0.80.
Key wordscontrast-enhanced ultrasound    time intensity curve    fitting analysis    parameter perfusion imaging    S-G filter
收稿日期: 2017-06-24     
PACS:  R318  
基金资助:国家自然科学基金 (81571675)
通讯作者: E-mail: ding.hong@zs-hospital.sh.cn   
引用本文:   
王本刚, 丁红, 彭诗云, 许智婷, 付甜甜, 王文平. 基于S-G滤波的超声造影分析参数灌注成像[J]. 中国生物医学工程学报, 2018, 37(4): 429-434.
Wang Bengang, Ding Hong, Peng Shiyun, Xu Zhiting, Fu Tiantian, Wang Wenping. Analysis of Parameter Perfusion Imaging for Contrast-Enhanced Ultrasound Based on S-G Filter. Chinese Journal of Biomedical Engineering, 2018, 37(4): 429-434.