|
|
Study on Imaging Effect of Different Excitation Patterns under Practical Electrical Impedance Tomography Conditions |
1 Institute of Biomedical Engineering, Chinese Academy of Medical Sciences & Peking Union Medical College, Tianjin 300192, China
2 No.291 Hospital of People〖DK〗’s Liberation Army, Baotou 014040, Inner Mongolia, China |
|
|
Abstract Using the commonly used adjacent excitation patternadjacent measurement pattern in the electrical impedance tomography (EIT), the practical detection system with finite measurement resolution (MR) and signaltonoise ratio (SNR) is always limited to distinguish tiny potential difference of electrodes and then finally make a great impact on image reconstruction. Based on the homogeneous field EIT boundary voltages analysis of adjacent excitation pattern, interval 6 electrodes and opposite excitation pattern in 16 electrodes array system, the MR and SNR requirements for imaging by practical EIT systems were analyzed, images reconstruction of three excitation patterns under different MR and different SNR conditions were simulated for three different image models A, B, C, which are positioned near to the center of the field, at the 1/2 radius of the field and near to the edge of the filed respectively. The combined TikhonovNoser regularization algorithm was adopted and image reconstruction errors function and structure similarity degree function for medical images quality evaluation were introduced to quantitative analysis the imaging effect. Imaging and evaluation results show that different excitation patterns have distinct index requirements of system MR and SNR. When MR was 1 mV, the imaging effect of interval 6 electrodes excitation pattern was the best, while adjacent excitation pattern was the best when MR was 0.01 mV. When MR was 0.01 mV, interval 6 electrodes excitation pattern was better than the adjacent excitation pattern for model A and B while the circumstance is opposite for model C. The critical value requirements of system SNR for Model A, B, C were 50 dB, 40 dB, 30 dB respectively under the interval 6 electrodes and opposite excitation patterns, which were 10 dB lower than the adjacent excitation pattern when imaging for all three models. The best imaging effects of all three model were under interval 6 electrodes excitation pattern when these SNR values near to critical values, followed by opposite excitation patterns. However, once the value of SNR was higher than the critical value, the best quality image was obtained under the adjacent excitation pattern. The main indexes impacting imaging results in low MR system and high MR system were different between the second smallest voltage and the smallest voltage and independent measurement number. It is advised to take interval 6 electrodes excitation pattern as the priority selection and opposite excitation pattern as the second one when system MR is relatively low, while choose adjacent excitation pattern when MR is high. For the 0.1 mV MR, interval 6 electrodes excitation pattern is suggested when imaging for model A and B and adjacent excitation pattern is suggested for model C. The factors impacting imaging results in low and high SNR system are the result of whole measurement voltage data and independent measurement number. When values of the system SNR are near to the critical values of 50 dB, 40 dB and 30 dB for model A, B, C respectively, interval 6 electrodes and opposite excitation pattern are suggested, and the adjacent excitation pattern is suggested once SNR exceeds critical value 10 dB.
|
|
|
|
|
[1]任超世,李章勇,王妍,等. 电阻抗断层成像应用基础与临床应用的一些研究进展[J]. 中国生物医学工程学报, 2010,29(2):300-304.
[2]邓娟,陈素华,沙洪,等. 信噪比对不同EIT图像重建算法的影响研究及评价[J]. 中国生物医学工程学报,2012,31(6):807-815.
[3]邓娟,陶凌,沙洪,等.测量分辨率对不同EIT算法影响的仿真研究[J]. 军事医学,2013,37(5):385-389.
[4]Seagar AD, Barber DC, Brown BH. Theoretical limits to sensitivity and resolution in impedance imaging [J]. Clin Phys Physiol Meas, 1987, 8(SA): S13-S31.
[5]Seagar AD, Brown BH. Limitations in hardware design in impedance imaging [J]. Clin Phys Physiol Meas, 1987, 8(SA): S85-S90.
[6]Cheney M, Isaacson D. Distinguishability in impedance imaging [J]. IEEE Trans Biomed Eng, 1992, 39(8):852-860.
[7]Rafiei\|Naeini M, McCann H. Lownoise current excitation subsystem for medical EIT [J]. Physiol Meas, 2008, 29(6S): 173-184.
[8]Wang W, Brown BH, Leatilard AD, et al. Noise equalisation within EIT images [J]. Physiol Meas, 1995, 15(S2A):211-216.
[9]Kim DY, Wi H, Yoo PJ, et al. Performance evaluation of KHU Mark2 parallel multifrequency EIT system [J]. J Phys: Conf Ser, 2010, 224(1): 13-16.
[10]Yasin M, Bhm S, Gaggero P, et al. Evaluation of EIT system performance[J]. Physiol Meas, 2011, 32(7):851-857.
[11]Adler A, Gaggero P, Maimaitijiang YS. Adjacent stimulation and measurement patterns considered harmful [J]. Physiol Meas, 2011, 32(S01):731-744.
[12]Adler A, Gaggero P, Maimaitijiang YS. Distinguishability in EIT using a hypothesistesting model [J]. J Phys: Conf Ser, 2010, 224(1):56-59.
[13]Avis NJ, Barber DC. Adjacent or polar drive: image reconstruction implications in electrical impedance tomography systems employing filtered back projection [M]. Paris: IEEE Press, 1992:87-88.
[14]Eyüboglu BM. An interleaved drive electrical impedance tomography image reconstruction algorithm [J]. Physiol Meas, 1996, 17(A):59-71.
[15]Franciane CP, Luis PSB. Some numerical results on the inuence of measurement strategies and load patterns in the EIT inverse problem [J]. J Phys: Conf Ser, 2010, 224(1):145-148.
[16]罗辞勇,陈民铀,王平. 电阻抗成像交叉测量模式的抗噪性能研究[J]. 仪器仪表学报,2009,1(1): 1-6.
[17]Lee EJ, Ts ME, Seo JK, et al. Breast EIT using a new projected image reconstruction method with multifrequency measurements [J]. Physiol Meas, 2012, 33(5):751-758.
[18]Grychtol B, Adler A,Choice of reconstructed tissue properties affects interpretation of lung EIT images[J]. Physiol Meas, 2014, 35(6):1035-1040.
[19]Zhao Zhanqi, Frerichs I, Pulletz S, et al. The influence of image reconstruction algorithms on linear thorax EIT image analysis of ventilation [J]. Physiol Meas, 2014, 〖STHZ〗35〖STBZ〗(6):1083-1089.
[20]Gaggero P, Adler A, Brunner J, et al. Electrical impedance tomography system based on active electrodes [J]. Physiol Meas, 2012, 33(5):831-847.
[21]Cherenin VA, Gulyaev YV, Korjenevsky AV, et al. An electrical impedance tomography system for gynecological application GIT with a tiny electrode array [J]. Physiol Meas, 2012, 33(5):849-862.
[22]Kao T J, Isaacson D, Newell JC, et al. A 3D reconstruction algorithm for EIT using a handheld probe for breast cancer detection [J]. Physiol Meas, 2006, 27(5):1-11.
[23]段影影,马建华,陈武凡,等. 改进的结构相似医学图像质量评价方法[J]. 计算机工程与应用,2010,46(2):145-149.
[24]Wang Zhou, Bovik AC, Sheikh HR, et al. Image Quality Assessment: From Error Measurement to Structural Similarity [J]. IEEE Transactions on Image Processing, 2004, 13(4):600-612.
[25]任超世,王妍. 阻抗成像实用化应用技术研究[J]. 中国医学装备,2007,4(5):1-5.
[26]Kolehmainen V, Voutilainen A, Kaipio JP. Estimation of nonstationary region boundaries in EITstate estimation approach [J]. Inverse Problems, 2001, 17(1):1937-1956.
[27]Wang Yan, Sha Hong, Ren Chaoshi. Optimum design of electrode structure and parameters in electrical impedance tomography [J]. Physiol Meas, 2006, 27(3):291-306.
|
|
|
|