Hybrid MOM/FDTD Electromagnetic Analysis Method and Application in MR RF Coil Design
School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China
Download: PDF (1069 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  The purpose of this paper is to improve the simulation accuracy of MR RF coil, then to improve the quality of MR image. We take advantages of the method of moment (MOM) and finite difference time domain (FDTD) method, that is, MOM is suitable for the calculation of current distribution on the RF coil structure, FDTD is suitable for the electromagnetic field analysis in the case of human body. We used the Huygens equivalent surface to combine the MOM and FDTD to form a new hybrid MOM/FDTD method. With the proposed real human body 3D electromagnetic model, we considered the complex coil tissue interactions effectively. The signal-to-noise ratio (SNR) of the image, which is designed by the method, was 193.4 dB, 38.7dB more than which does not consider the interactions. Consequently, the results verified the accuracy of the design.

Key wordsmethod of moment (MOM)      finite difference time domain (FDTD)      Huygens equivalent surface      magnetic resonance radiation frequency coil (MR RF coil)
     
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
WANG Di HUANG Qi Hua XIN Xue Gang*
Cite this article:   
WANG Di HUANG Qi Hua XIN Xue Gang*. Hybrid MOM/FDTD Electromagnetic Analysis Method and Application in MR RF Coil Design[J]. journal1, 2012, 31(2): 262-267.
URL:  
http://cjbme.csbme.org/EN/ 10.3969/j.issn.0258-8021.2012.02.015     OR     http://cjbme.csbme.org/EN/Y2012/V31/I2/262
Copyright © Editorial Board of Chinese Journal of Biomedical Engineering
Supported by:Beijing Magtech