1 National Regional Key Technology Engineering Laboratory for Medical Ultrasound,Department of Biomedical Engineering, School of Medicine, Shenzhen University, Shenzhen 518060, Guangdong, China 2Shenzhen Mindray Bio-Medical Electronics Co.,LTD, Shenzhen 518057, Guangdong, China
Abstract:Diabetes is becoming one of the highest incidence diseases of China. The traditional glucose self-monitor need to take drops of blood from the fingers of the patient. This method is painful and also has the possibility of infection. Thus, in this review, we introduced the principles and development of non-invasive glucose monitor, and formed a research frame of non-invasive glucose monitor.
黄絮, 李勇, 陈思平. 无创及微创血糖仪研究进展[J]. 中国生物医学工程学报, 2016, 35(5): 598-604.
Huang Xu Li Yong Chen Siping. The Development of Non-Invasive Glucose Monitor. Chinese Journal of Biomedical Engineering, 2016, 35(5): 598-604.
[1] International Diabetes Federation[R]. Annual Report 2014. www.idf.org. [2] Yang WY, Lu JM, Weng JP, et al. Prevalence of diabetes among men and women in China[J]. N Engl J Med, 2010,362(25):2425-2425. [3] Xu Y, Wang LM, He J, et al. Prevalence and control of diabetes in Chinese adults[J]. JAMA, 2013,310(9):948-959. [4] Porte D, Schwartz MW. Diabetes complications: why is glucose potentially toxic?[J]. Science,1996,272(5262):699-700. [5] Georgia P, Tina C, Demetrius E, et al. The 30-year natural history of type 1 diabetes complications: the Pittsburgh Epidemiology of Diabetes Complications Study experience[J]. Diabetes, 2006,55(5):1463-1469. [6] Perkins BA, Aiello LP, Krolewski AS. Diabetes complications and the renin-angiotensin system[J]. N Engl J Med,2009,361(1):83-85. [7] Intine RV, Sarras MP. Metabolic memory and chronic diabetes complications: Potential role for epigenetic mechanisms[J]. Curr Diab Rep, 2012,12(5):551-559. [8] Buckloh LM, Lochrie AS, Holly A, et al. Diabetes complications in youth[J]. Diabetes Care, 2008,31(8):1516-1520. [9] MP Gilbert. Screening and treatment by the primary care provider of common diabetes complications[J]. Med Clin North Am, 2015,99(1):201-219. [10] Shah BR, Hwee J, Cauch-Dudek K, et al. Diabetes self-management education is not associated with a reduction in long-term diabetes complications: an effectiveness study in an elderly population[J]. Journal Eval in Clin Pract,2015,21(4):656-661. [11] Kuhn M, Kuhn P, Villeval MC. Self control and inter temporal choice: evidence from glucose and depletion interventions[C]//CESifo Working Paper Series. Munich: CES, 2014: No 4609. [12] Beedie CJ,Lane AM. The role of glucose in self-control: Another look at the evidence and an alternative conceptualization[J]. Pers Soc Psychol Rev, 2011,16(2):143-153. [13] Zamzow KL, Schmidt DG, Shults MC, et al. Development and evaluation of a wearable blood glucose monitor[J]. Assaio Transactions,1990,36(3):M588-M591. [14] Rayman G, Spencer PD, Tillyer CR, et al. Evaluation of a self-calibrating blood glucose monitor[J]. Diabetes Care, 1984,7(4):378-380. [15] Jones PM. Use of a course on self-control behavior techniques to increase adherence to prescribed frequency for self-monitoring blood glucose[J]. Diabetes Educ,1990,16(4):296-303. [16] Fine J. Glucose content of normal urine [J]. British Medical Journal. 1965,1(5444):1209-1214. [17] Street HV. Determination of glucose in biological fluids with ethylenediaminetetra-acetic acid[J]. Analyst,1958,83(992):628-634. [18] Bantle JP, Thomas W. Glucose measurement in patients with diabetes mellitus with dermal interstitial fluid[J]. J Lab Clinl Med,1997,130(4):436-441. [19] Vesper HW, Wang PM, Archibold E, et al. Assessment of trueness of a glucose monitor using interstitial fluid and whole blood as specimen matrix[J]. Diabetes Technol Ther,2006,8(1):76-80. [20] Rebrin K, Sheppard NF, Steil GM. Use of subcutaneous interstitial fluid glucose to estimate blood glucose: revisiting delay and sensor offset[J]. J Diabetes Sci Technol,2010, 4(5):1087-1098. [21] Weiss TR. Google Testing Smart Contact Lenses for Diabetes Patients[N]. Eweek,2014. [22] Zhang J, Hodge W, Hutnick C, et al. Noninvasive diagnostic devices for diabetes through measuring tear glucose[J].J Diabetes Sci Technol,2011,5(1):166-172. [23] Baca JT, Finegold DN, Asher SA. Tear Glucose Analysis for the noninvasive detection and monitoring of diabetes mellitus[J]. Ocul Surf,2007,5(4)280-293. [24] Yao HF, Shum AJ, Cowan M. et.al. A contact lens with embedded sensor for monitoring tear glucose level[J]. Biosens Bioelectron, 2011. 15:26(7): 3290-3296. [25] Yang X, Pan X, Blyth J, et al. Towards the real-time monitoring of glucose in tear fluid: holographic glucose sensors with reduced interference from lactate and pH[J]. Biosens Bioelectron, 2008,23(6):899-905. [26] Lee H, Choi TK, Lee YB, et al. A graphene-based electrochemical device with thermoresponsive microneedles for diabetes monitoring and therapy[J]. Nat Nanotechnol, 2016,11(6):566-572. [27] Cengiz E, Tamborlane WV. A tale of two compartments: interstitial versus blood glucose monitoring[J]. Diabetes Technol Ther, 2009,11(Suppl 1):S11-S16. [28] Mraovic B. Analysis: Continuous glucose monitoring during intensive insulin therapy[J]. Diabetes Technol Ther, 2009.3(4):960-963. [29] Buckingham B, Caswell K, Wilson DM. Real-time continuous glucose monitoring[J]. Curr Opin Endocrinol Diabetes Obes, 2007,14(4):288-295. [30] Taylor JR. What′s new in continuous glucose monitors?[J]. Endocrine Today,2009(8):1-2. [31] Hermanns N, Schumann B, KulzerB, et al. The impact of continuous Glucose monitoring on low interstitial glucose values and low blood glucose values assessed by point-of-care blood glucose meters: Results of a crossover trial[J]. J Diabetes Technol Ther, 2014.8(3):516-522. [32] Facchinetti A, Del Favero S, Sparacino G. et al. Model of glucose sensor error components: identification and assessment for new Dexcom G4 generation devices[J]. Med Biol Eng Comput, 2015, 53(12):1-11 [33] Facchinetti A, Del Favero S, Sparacino G. et al. Modeling the glucose sensor error[J].IEEE Trans Biomed Eng, 2014,61(3):620-629. [34] Peyser TA, Nakamura K, Price D, et al. Hypoglycemic accuracy and improved low glucose alerts of the latest Dexcom G4 platinum continuous glucose monitoring system[J]. Diabetes Technol Ther, 2015,17(8):548-554. [35] Hirsch IB. Clinical review: Realistic expectations and practical use of continuous glucose monitoring for the endocrinologist[J]. J Clin Endocrinol Metab,2009;94(7):2232-2238 [36] Dassau E, Cameron F, Lee H, et al. Real-time hypoglycemia prediction suite using continuous glucose monitoring: a safety net for the artificial pancreas[J]. Diabetes Care,2010,33(6):1249-1254. [37] Turner A, Gifford R. Implanted sensors[J]. Springer Series on Chemical Sensors and Biosensors, 2012.13:159-189. [38] Tierney MJ, Tamada JA, Potts RO, et al. The GlucoWatch biographer: a frequent automatic and noninvasive glucose monitor[J].Ann Med, 2000, 32(9):632-641. [39] Keenan DB, Mastrototaro JJ, Voskanyan G, et al. Delays in minimally invasive continuous glucose monitoring devices: a review of current technology[J]. J Diabetes Sci Technol, 2009, 3(5):1207-1214. [40] Mazze RS, Strock E, Borgman S, et.al. Evaluating the accuracy, reliability, and clinical applicability of continuous glucose monitoring(CGM):Is CGM ready for real time?[J]. Diabetes Technol Ther, 2009,11(1):11-18. [41] Fischbacher C, Jagemann KU, Danzer K, et al. Enhancing calibration models for non-invasive near-Infrared spectroscopicalblood glucose determination[J]. Fresen J Anal Chem,1997, 359(1):78-82. [42] Liakat S, Bors KA, Huang TY, et al. In vitro measurements of physiological glucose concentration in biological fluids using mid-infrared light[J]. Biomed Opt Express, 2013,4(7):1083-1090. [43] Cooper PJ, Barker TQ. Individual Calibration of blood glucose for supporting noninvasive self-monitoring blood glucose[P]. EP, US6309884 B1. 2001. [44] Barman I. Unraveling the puzzles of spectroscopy-based non-invasive blood glucose detection[D]. Boston: Massachusetts Institute of Technology, 2011. [45] Liakat S, Bors KA, Xu L, et al. Noninvasive in vivo glucose sensing on human subjects using mid-infrared light[J]. Biomed Opt Express, 2014,5(7):2397-2404. [46] Gao L,Tao W,Zhao H,et al. Research and design of photoacoustic cell of noninvasive blood glucose measurement[J]. Optical Instruments, 2012,34(2):55-60. [47] Xing Zhang, Li Xiaoqing. Strictly regulated blood glucose oscillations linked to rat mortality after severe burn injury[J]. Faseb J, 2012, 26(9):832-842. [48] Zhao Z, Myllyla RA. Photoacoustic blood glucose and skin measurement based on optical scattering effect[J]. Proceedings of SPIE,2002,4707:1-5. [49] Ren Z, Liu G, Huang Z, et al. Laser-induced photoacoustic glucose spectrum denoising using an improved wavelet threshold translation-invariant algorithm[J]. Proceedings of SPIE, 2009, 7382:1-8. [50] Firdous S, Nawaz M, Ahmed M, et al. Measurement of diabetic sugar concentration in human blood using Raman spectroscopy[J]. Laser Physics, 2012, 22(6):1090-1094 [51] Dingari NC, Barman I, Singh GP, et al. Investigation of the specificity of Raman spectroscopy in non-invasive blood glucose measurements[J]. Anal Bioanal Chem,2011,400(9):2871-2880 [52] Enejder AM, Koo TW, Oh J, et al. Blood analysis by Raman spectroscopy[J]. Optics Lett, 2002, 27(22):2004-2006. [53] Shao J, Lin M, Li Y, et al. In vivo blood glucose quantification using raman spectroscopy[J]. PLoS ONE, 2012, 7(10):e48127 [54] Segman Y. Finger mounted non-invasive device for measuring physiological parameters[P]. US687150. 2013 [55] Segman Y. Finger deployed device for measuring blood and physiological characteristics[P]. US8489165. 2013 [56] Lipson J, Bernhardt J, Block U, et al. Non-invasive technologies for glucose monitoring: requirements for calibration in noninvasive glucose monitoring by Raman spectroscopy[J]. J Diabetes Sci Technol, 2009,3(2):233-241. [57] Lipson J,Bernhardt J,U Block, et al. Requirements for calibration in noninvasive glucose monitoring by Raman spectroscopy[J]. J Diabetes Sci Technol, 2009,3(2): 233-241. [58] Harman\|BoehmI I, Gal A, Raykhman AM, et al. Noninvasive glucose monitoring: Increasing accuracy by combination of multi-technology and multi-sensors[J]. J Diabetes Sci Technol, 2010,4(3):583-595. [59] Harman-Boehm I, Gal A, Raykhman AM, et al. Noninvasive glucose monitoring: a novel approach[J]. J Diabetes Sci Technol, 2009,3(2):253-60. [60] Ramchandani N, Heptulla RA. New technologies for diabetes: a review of the present and the future[J]. Int J of Pediatric Endocrinol,2012,2012(1):1-10. [61] Gravesen P, Raaby PK, Dirac H. Lab-on-a-chip technology for continuous glucose monitoring[J]. J Diabetes Sci Technol,2007, 1(3):372-374. [62] Pedersen A. GlucoTrack offers painless glucose-monitoring option[J]. Medical Device Daily, 2013, 17(197): 1-7.