Abstract:This study proposed a three-type classification method for wearable ECG signal quality assessment. Rationale for three-type classification is from the clinical requirement for diagnosis,and the three types for signal quality are:1) clinically useful signals with good signal quality;2) clinically useful signals with poor signal quality;3) clinically useless signals. The method firstly extracted 12 signal quality features from time-domain,frequency-domain and nonlinear-domain,and constructed the feature matrix. Then a support vector machine (SVM) classifier based on a radial basis kernel function was trained on the collected wearable ECG signals with manual annotations. Results from 375 independent test ECG samples with clinical labels showed that the proposed method achieved an F measure of 0.909,0.827 and 0.973 for the three-type quality signals respectively,and the overall classification accuracy was 92.3%,which was 2.2% and 6.4% higher than the comparable methods,i.e.,the CNN-based model and the traditional SVM model. This study demonstrated that the new three-type classification model of signal quality had application potentials in wearable dynamic ECG signal quality assessment.
[1] 刘澄玉,杨美程,邸佳楠,等.穿戴式心电:发展历程、核心技术与未来挑战[J].中国生物医学工程学报,2019,38(6):641-652. [2] Monasterio V,Burgess F,Clifford GD.Robust classification of neonatal apnoea-related desaturations[J].Physiol Meas,2012,33(9):1503-1516. [3] Redmond SJ.Electrocardiogram signal quality measures for unsupervised telehealth environments[J].Physiol Meas,2012,33(9):1517-1533. [4] Xia Henian,Garcia GA,Bains J,et al.Matrix of regularity for improving the quality of ECGs[J].Physiol Meas,2012,33(9):1535-1548. [5] Hayn D,Jammerbund B,Schreier G.ECG quality assessment for patient empowerment in mHealth applications[C]//2011 Computing in Cardiology.Hangzhou:IEEE,2011:353-356. [6] Langley P,King S.An algorithm for assessment of quality of ECGs acquired via mobile telephones[C]//2011 Computing in Cardiology.Hangzhou:IEEE,2011:281-284. [7] Moody BE.Rule-based methods for ECG quality control[C]//2011 Computing in Cardiology.Hangzhou:IEEE,2011:361-363. [8] Chudáček V,Zach L,Kuílek J,et al.Simple scoring system for ECG quality assessment on Android platform[C]//2011 Computing in Cardiology.Hangzhou:IEEE,2011:449-451. [9] Xia Henian,Garcia GA,McBride JC,et al.Computer algorithms for evaluating the quality of ECGs in real time[C]//2011 Computing in Cardiology.Hangzhou:IEEE,2011:369-372. [10] 李延军,唐晓英,许志,等.一种基于波形特征的12导联心电信号质量的估计方法[J].航天医学与医学工程,2015,28(6):419-426. [11] Kalkstein N,Kinar Y,Na′aman M,et al.Using machine learning to detect problems in ECG data collection[C]//2011 Computing in Cardiology.Hangzhou:IEEE,2011:437-440. [12] Zaunseder S,Huhle R,Malberg H.CinC challenge — Assessing the usability of ECG by ensemble decision trees[C]//2011 Computing in Cardiology.Hangzhou:IEEE,2011:277-280. [13] 张亚涛.移动心电信号质量评估的分析与研究[D].济南:山东大学,2015. [14] Zhang Yatao,Liu Chengyu,Wei Shoushui,et al.ECG quality assessment based on a kernel support vector machine and genetic algorithm with a feature matrix[J].Journal of Zhejiang University SCIENCE C,2014,15(7):564-573. [15] 寇明春.心电信号质量评估方法研究[D].兰州:兰州理工大学,2014. [16] Li Qiao,Mark RG,Clifford GD.Robust heart rate estimation from multiple asynchronous noisy sources using signal quality indices and a Kalman filter[J].Physiol Meas,2008,29(1):15-32. [17] 李桥,俞梦孙.重症监护病人心电导联信号质量评估[J].山东大学学报(医学版),2007,45(9):868-872. [18] Thakor NV,Webster JG,Tompkins WJ.Estimation of QRS complex power spectra for design of a QRS filter[J].IEEE Trans Biomed Eng,1984,31(11):702-706. [19] Hamilton PS,Tompkins WJ.Quantitative investigation of QRS detection rules using the MIT/BIH arrhythmia database[J].IEEE Trans Biomed Eng,1986,33(12):1157-1165. [20] Manikandana MS,Soman KP.A novel method for detecting R-peaks in electrocardiogram (ECG) signal[J].Biomed Signal Process Control,2012,7(2):118-128. [21] Richman JS,Moorman JR.Physiological time-series analysis using approximate entropy and sample entropy[J].Am J Physiol Heart Circ Physiol,2000,278(6):H2039-H2049. [22] Zhao Zhongyao,Liu Chengyu,Li Yixuan,et al.Noise rejection for wearable ECGs using modified frequency slice wavelet transform and convolutional neural networks[J].IEEE Access,2019,7:34060-34067. [23] Behar J.Oster J,Li Qiao,et al.A single channel ECG quality metric[C]//2012 Computing in Cardiology.Krakow:IEEE,2012:381-384.